Ac Power Box (Fru No. 704) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-9-5

AC Power Box (FRU No. 704)

Purpose: This is a description on how to remove and replace the AC Power Box.
8-9-5-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
8-9-5-2
Needed Manpower
1 person, 15 minutes + travel
8-9-5-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
AC Power Assy should be removed before proceed. Refer to AC POWER ASSY DISASSEMBLY
on page 8 - 82.
8-9-5-4
Removal Procedure
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
CAUTION
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
1.) Unscrew six screws (1-6) and remove the AC Power Box cover. Refer to Figure 8-182.
2.) Remove the AC Power left side cover.
7
AC Power Box Cover
Figure 8-182 Unscrew 6 screws to remove AC Power Box
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 163

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents