Section 8-10

Software Loading Procedure

8-10-1
General
This describes a full system software loading (Base Software Load Image + LOGIQ 5 PRO Application)
procedure.
Use this instruction when:
- Performing the HDD with new one (Base Software Load Image + Application)
- Performing the system revision-up (Application)
Reference : The below is the support table between Application Software version and Base System
Software Load Image part number
Table 8-55
Application
Software Version
R4.X.X
8-10-2
Parts Required
Base System Software Load Image CD
LOGIQ 5 PRO R4.x.x Application Software
Blank CD-R ( for patient image backup)
Service Dongle
NOTE:
Patient Image Data will be deleted when all the drives are ghosted. Back up them on the CD-R
or MOD before starting software loading
NOTE:
Insert the service dongle into ths service port located at the rear panel before installing the Base
System Software Load image
8-10-3
TimeRequired
Approximately 2 hour
8 - 166
7
Base Software Load Image CD Part Numbers
Base Software Load Image Part P/N
5134559-x (BEP2)
Supported