Gems Leakage Current Limits - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
10-7-2

GEMS Leakage Current Limits

The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment
Safety Standards. These limits are GEMS standards and in some cases are lower than the above
standards listed.
Table 10-12 Chassis Leakage Current Limits—Accessible Metal Surfaces
Country
Other
Table 10-13 Type BF Applied Part Leakage Current Limits - Non-Conductive (Floating) Surface
Country
USA
Other
Table 10-14 Type CF Applied Part Leakage Current Limits - Surgical Probes and ECG
Country
USA
Other
NOTE:
*Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the
part to determine the amount of current that will pass (or sink) to ground if a patient contacted
mains voltage.
The following tests are performed at the factory and should be performed at the site. These tests are:
grounding continuity, chassis leakage current, probe leakage current, and ECG leakage current. All
measurements are made with an electrical safety analyzer.
7
Normal Condition
N/A
USA
0.1 mA
and Cavity Probes
Normal
Condition
0.05 mA
0.1 mA
Connections
Normal
Condition
0.01 mA
0.01 mA
Chapter 10 Care & Maintenance
Open Ground
0.3 mA
0.5 mA
Open Ground
Reverse Polarity
0.05 mA
0.05 mA
0.5 mA
0.5 mA
Open Ground
Reverse Polarity
0.05mA
0.05 mA
0.05 mA
0.05 mA
LOGIQ™5 PRO S
ERVICE
Reverse Polarity
Open Neutral
0.3 mA
0.5 mA
Open Neutral
*Mains Applied
0.05 mA
0.5 mA
Open Neutral
*Mains Applied
N/A
0.05 mA
M
ANUAL
N/A
0.5 mA
N/A
5.0 mA
0.025 mA
0.05 mA
10-11

Hide quick links: