Chapter 10
Care & Maintenance
Section 10-1

Overview

10-1-1
Periodic Maintenance Inspections
It has been determined by engineering that your LOGIQ™ 5 system does not have any high wear
components that fail with use, therefore no Periodic Maintenance Inspections are mandatory. Some
Customers Quality Assurance Programs may require additional tasks and or inspections at a different
frequency than listed in this manual.
10-1-2
Purpose of Chapter 10
This chapter describes Care & Maintenance on the scanner and peripherals. These procedures are
intended to maintain the quality of the ultrasound systems performance. Read this chapter
completely and familiarize yourself with the procedures before performing a task.
Table 10-1
Section
10-1
10-2
10-3
10-4
10-5
10-6
10-7
10-8
Practice good ESD prevention. Wear an anti–static strap when handling electronic parts and
CAUTION
even when disconnecting/connecting cables.
THERE ARE SEVERAL PLACES ON THE BACKPLANE, THE AC DISTRIBUTION,
DANGER
AND DC DISTRIBUTION THAT ARE DANGEROUS. BE SURE TO DISCONNECT
THE SYSTEM POWER PLUG AND OPEN THE MAIN CIRCUIT BREAKER
BEFORE YOU REMOVE ANY PARTS. BE CAUTIOUS WHENEVER POWER IS
STILL ON AND COVERS ARE REMOVED.
Do not pull out or insert circuit boards while power is ON.
CAUTION
7
Contents in Chapter 10
Overview
Why do Maintenance
Maintenance Task Schedule
Tools Required
System Maintenance
Using a Phantom
Electrical Safety Tests
When There's Too Much Leakage Current...
Ultrasound INSPECTION Certificate
Chapter 10 Care & Maintenance
LOGIQ™5 PRO S
Description
M
ERVICE
ANUAL
Page Number
10-1
10-2
10-2
10-4
10-5
10-10
10-10
10-22
10-23
10-1