Monitor Disassembly And Reassembly; Crt Assy (Fru No. 100) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
Section 8-3
Monitor
8-3-1

CRT Assy (FRU No. 100)

Purpose: This is a description on how to remove and replace the CRT Assy.
8-3-1-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
8-3-1-2
Needed Manpower
2 persons, 5 minutes + travel
8-3-1-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Maneuver Control Console to a suitable position for removing the monitor.
8-3-1-4
Removal procedure
1.) Remove the OP Rear Cover. Refer to section 8-5-10 on page 58.
2.) Unscrew one screw cap located at (1).
8 - 2
7
Figure 8-1 Removing OP rear Cover and Unscrew screw
Section 8-3 - Monitor