Op Panel Lcd Assy Validation - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
4-11-8

OP Panel LCD Assy Validation

1.) Boot up the system.
2.) Follow up the process from step1 to step7 of Section 4-11-1.
3.) Touch the touchpanel.
In case of touch point is displayed in log of scfpapitest.exe while you press the touchpanel, LCD assy
NOTICE
is connected correctly.
7
Figure 4-47 Touchpanel
Chapter 4 Functional Checks
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
4 - 33