Speaker Assy (FRU No. 104)

Purpose: This is a description on how to remove and replace the Speaker Assy.
8-3-5-1
Tools
Common pillips screwdrivers
8-3-5-2
Needed Manpower
1person, 10 minutes + travel
8-3-5-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Monitor Cover Set should be removed before proceed. For more information Refer to section 8-3-
3 on page 8.
8-3-5-4
Removal Procedure
1.) Disconnect two connectors (1-2) from the rear side of the Monitor. Refer to Figure 8-13.
8 - 12
7
2
Figure 8-13 Disconnect two connectors