Data Sheet For Transducer Source Leakage Current - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
10-7-9-4

Data Sheet for Transducer Source Leakage Current

The test passes when all readings measure less than the values shown in
14. Record all data on the PM Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
CAUTION
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged
.
Table 10-21 Typical Data Sheet For Transducer Source Leakage Current
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
7
Tester Power Polarity
Switch
NORM
NORM
REV
REV
NORM
NORM
REV
REV
Chapter 10 Care & Maintenance
Transducer Tested:
Tester GROUND or
NEUTRAL Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
LOGIQ™5 PRO S
ERVICE
Table 10-13
and
Table 10-
Measurement
M
ANUAL
10-21