Front Cover (FRU No. 306)

Purpose: This is a description on how to remove and replace the Front Cover.
8-5-7-1
Tools
Common pillips screwdrivers
8-5-7-2
Needed Manpower
1person, 20 minutes + travel
8-5-7-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-5-7-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Front Base Cover (FRU No.306). Refer to section 8-5-6 on page 51.
4.) Unscrew four screws (1-4). Refer to Figure 8-51.
(2)
5.) Remove the Front Cover.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-26
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-12-1
Section 10-5-5
8-5-7-5
Mounting procedure
Install the new parts in the reverse order of removal.
7
Figure 8-51 Unscrew 2 screws on the left side
Functional Tests
Section
Functional Test / Diagnostic Test
Cover Parts Function Validation
Chapter 8 Replacement Procedures
Left Side
Power On/Boot Up
Power Off / Shutdown
Physical Inspection
LOGIQ™ 5 PRO S
(1)
(3),(4) are on the right side
Debrief Script
"Service Manual, Direction
2380207
, Rev 7+, Section 8-5-7. Equipment
passed all required tests and is ready for use. "