SMPS, HDD, ODD Assy function check procedure

4-13-13-1
SMPS Assy Test
1.) Turn on the system and check the completion of Echoloader loading.
2.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4-
7. Refer to Section 4-3-5 - System CFM and PWD Checks on page 4-8.
4-13-13-2
HDD Test
1.) Refer to 4-13-11-2 - Basic Function Test.
4-13-13-3
ODD Test
1.) Turn on the system and check the completion of Echoloader loading.
2.) Enter "Patient" Mode
3.) Select "Transfer" button and then Insert CD/DVD media which is included patient data. Execute
"Import".
Check display 'Success' message.
4.) Select "Transfer" button and then execute "Export". Insert empty CD/DVD media.
Check display 'Success' message.
5.) Execute "Ez back up". Insert empty CD/DVD media.
Check display 'Success' message.