CWDOP Assy function check procedure
1.) Turn on the system and check the completion of Echoloader loading.
2.) Do the Service Login. Refer to the Section 5-10-2-2 - Service Login on page 5-42.
3.) Select 'Diagnostics'=>'LOGIQ 5 Diagnostics'=>'FRU Test'=>'CWDOP Full Test'
4.) Execute the diagnostic test.
5.) All diagnostic test item of CWDOP should be passed.
6.) Reboot up the system
7.) Connect Probe in Probe port (Sector or Pencil Probe).
8.) After system boot up, select the CWD mode.
9.) Check the system basic functions. Refer to Section 4-3-5 - System CFM and PWD Checks on page
4-8.
4 - 50
7
Figure 4-67 CWDOP Diagnostic test