Op Side R Cover (Fru No. 308) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-5-9

OP Side R Cover (FRU No. 308)

Purpose: This is a description on how to remove and replace the OP Side R Cover.
8-5-9-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
8-5-9-2
Needed Manpower
1person, 25 minutes + travel
8-5-9-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-5-9-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Front Base Cover (FRU No.306). Refer to 1.) on page 8-40.
4.) Remove the Front Cover (FRU No.306). Refer to section 8-5-7 on page 53.
5.) Remove the Keyboard Assy (FRU No,200). Refer to section 8-4-1 on page 16.
6.) Unscrew five (5) screws (1-5) and remove the Probe Holder Bracket. Figure 8-54.
8 - 56
7
(1)
Probe Holder Bracket
Figure 8-54 Removing the Probe Holder Bracket
Section 8-5 - Covers
(4) (5)
(3)
(2)
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL

Hide quick links: