Transformer (Fru No. 703) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-9-4

Transformer (FRU No. 703)

Purpose: This is a description on how to remove and replace the Transformer.
8-9-4-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
8-9-4-2
Needed Manpower
1 person, 15 minutes + travel
8-9-4-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-9-4-4
Removal Procedure
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
CAUTION
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
1.) Remove the AC Power top cover. Refer to Figure 8-83 on page 86 .
2.) Remove the AC Power left cover. Refer to Figure 8-82 on page 84 .
3.) Disconnect four (4) connectors (1-4, J3, J6, J9, J11). Refer to Figure 8-179..
4.) Cut the Cable ties tied to the cables.
5.) Disassemble the AC Power box cover. Refer to Figure 8-85 on page 90 .
7
(1)
(2)
J9
J11
JPC Board
Figure 8-179 Disconnecting 4 connectors
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
(4)
(3)
J6
J3
M
ERVICE
ANUAL
8 - 161

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents