Op Panel Tgc Assy (Fru No. 205) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-4-6

OP Panel TGC Assy (FRU No. 205)

Purpose: This is a description on how to remove and replace OP Panel TGC Assy.
8-4-6-1
Tools
Common pillips screwdrivers.
8-4-6-2
Needed Manpower
1persons, 5 minutes + travel
8-4-6-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
page 16.
8-4-6-4
Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-27.
2.) Disconnect the TGC connector from the back side of the keyboard.
3.) Unscrew four (4) screws and Remove the TGC Assy from the keyboard.
7
Figure 8-27 Removing TGC Knobs
1
2
Figure 8-28 Unscrew 4 screws and remove TGC Assy
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
ERVICE
TGC Connector
3
4
M
ANUAL
8 - 27

Hide quick links: