Pc Diagnostics (Interactive Tests) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
7-6-3

PC Diagnostics (Interactive Tests)

7-6-3-1
Keyboard Test
7-6-3-2
Audio Test
7-6-3-3
Monitor Test
When you select a test to run, the display is changed.
13.)Instructions Frame
-
14.)Status Frame
-
-
15.)Execute Button
-
-
-
16.)Loop Count
-
17.)Progress Indicator
-
18.)Short Text Message
-
19.)Status Frame Background Color
-
-
-
-
7
Displays either a test-specific text or the default instructions.
Initially displays the last known status for a selected diagnostic. Once the diagnostic starts,
the frame displays the "current" status of all test results.
The Status Frame also contains the user interface elements used for Diagnostic Control and
Operator Feedback.
This switch has two modes - each with appropriate text:
Execute - to start the diagnostic
Abort - to stop a diagnostic
This is an editable text field that will only accept numeric values of 4 digits or less. When the
switch is configured as an "execute" switch and pressed, the loop count field will be queried
to determine the number of times to execute the diagnostic.
Displays a graphical progress indicator for the user.
Displays brief messages about the test's progress during execution.
Initially gray, the Status Frame background color changes upon completion of a diagnostic to
indicate completion status.
Code Status Fail = Red
Code Status Pass = Green
Code Status neither Pass nor Fail (Example: Aborted) = Gray (default color).