Power Off/ Shutdown

NOTE:
After turning off a system, wait at least ten seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
3-5-2-1
Power Shutdown
1.) Press once the ON/OFF key at the front of the System.
2.) Click the shutdown button in the Exit dialog box as Figure 3-17 on page 3-14
3.) Switch OFF the Main Circuit Breaker at the rear of the system Refer to Figure 3-13 on page 3-11 .
4.) Disconnect the Main Power Cable is necessary. For example : Relocating the scanner.
Disconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the
NOTICE
uninterruptible power system (UPS) in the backend processor, forcing an ordered shutdown of the
system.
The UPS serves only as a battery backup that allows for a soft shut-down of the scanner to prevent file
NOTICE
corruption. IT DOES NOT REGULATE INCOMING VOLTAGE.
3 - 14
7
Figure 3-17 System Exit Window