Scan Probe Environmental Requirements - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
Table 2-5
Replace all screws, RF
gaskets, covers, cores
Replace broken RF gaskets
Do not place labels where RF
gaskets touch metal
Use GE specified harnesses
and peripherals
Take care with cellular phones Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Properly dress peripheral
cables
2-2-4

Scan Probe Environmental Requirements

Operation:10° to 35° C
Storage:--10° to 60° C
NOTE:
Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
NOTICE
OF -20 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
7
EMI Prevention/abatement
EMI Rule
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end. Install the
shield over the front of card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interface with the ultrasound signals.
If more than 20% or a pair of fingers on the RF gaskets are broken, replace the gaskets.
Do not turn on the unit until any loose metallic part is removed.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. Or, if a label has been found in such a position, move the label.
The interconnect cables are grounded and require ferrite beads and other shielding. Also,
cable length, material, and routing are all important; do not change from what is specified.
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor
cables to the frame.
Chapter 2 Pre-Installation
LOGIQ™5 PRO S
Details
M
ERVICE
ANUAL
2-5

Hide quick links:

Advertisement

Table of Contents
loading