Task Lamp Assy Check; Keyboard Front Grip Check; Probe Holder Check - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
4-12-1-1
Task Lamp Assy
Check if task lamp is operated.
4-12-1-2
Keyboard Front Grip
Check if the keyboard front grip is fastened so that it won't move.
Check if the keyboard front grip can not be removed easily.
4-12-1-3
Probe holder
Check if the probe holder is fastened so that it won't move.
Check if the probe holder can not be removed easily.
4 - 40
7
Figure 4-53 Task Lamp Assy
Figure 4-54 Keyboard Front Grip
Figure 4-55 Probe holder
Section 4-12 - Mechanical Function Checks
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL