Section 4-10Monitor Function Checks
4-10-1

CRT Assy function check procedure

1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
Figure 4-40.
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.
4-10-2

Monitor Cable Assy function check procedure

1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
Figure 4-40.
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.
4-10-3

Monitor S/W Assy function check procedure

1.) Turn on the system and check the completion of Echoloader loading.
2.) Select "Utility Key" on the Key board and then choose " Test Pattern" and "Resolution" marked in
Figure 4-40.
3.) All the Numbers from '0' to '100' on the Monitor Test pattern screen above should be easily
distinguished.