Rear Caster (FRU No. 404)

Purpose: This is a description on how to remove and replace the Rear Caster.
8-6-4-1
Tools
Common pillips screwdrivers
Allen/Unbraco wrench
Lifter (180Kg)
8-6-4-2
Needed Manpower
1 person, 15 minutes + travel
8-6-4-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Lift the system before proceed.
8-6-4-4
Removal Procedure
1.) Unscrew eight hexagonal bolts (1-8).
2.) Remove the Rear Caster. Refer to Figure 8-76.
7
6
5
7
2
1
4
3
Figure 8-76 Removing the Rear Caster