Air Filter Set (Fru No. 406) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-6-6

Air Filter Set (FRU No. 406)

Purpose: This is a description on how to remove and replace the Air Filter Set.
8-6-6-1
Tools
Not Required.
8-6-6-2
Needed Manpower
1 person, 5 minutes + travel
8-6-6-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-6-6-4
Removal Procedure
1.) Remove Air Filter as illustrated below.Refer to Figure 8-78.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-41
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-12-1-7
Section 10-5-5
8-6-6-5
Mounting procedure
Install the new parts in the reverse order of removal.
7
Figure 8-78 Removing the Air Filters
Functional Tests
Section
Functional Test / Diagnostic Test
Chapter 8 Replacement Procedures
Power On/Boot Up
Power Off / Shutdown
Air filter set
Physical Inspection
LOGIQ™ 5 PRO S
ERVICE
Debrief Script
"Service Manual, Direction
2380207
, Rev 7+, Section 8-6-6. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8 - 87

Hide quick links: