Backup and Restore Database, Preset Configurations and Images

Avoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator
NOTICE
unlocked and the MOD susceptible to damage if moved. If forced to use this method, reboot the
system, then insert and eject a known good disk using one of the other manual ejection
methods.
4-3-11-1
Formatting Media
1.) To format the backup media, CD-ROM or MOD, select the UTILITY button on the Menu panel.
2.) Select CONNECTIVITY, then TOOL( in case of BT05 SW, REMOVABLE MEDIA )on the
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-31. It is best to use all capital letters with no
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-32. When the formatting
6.) If desired, verify that the format was successful by returning to Utility>Connectivity>Tools (in
4 - 16
7
monitor screen. Properly label and Insert the backup media.
spaces or punctuation marks.
Figure 4-31 Format and Verify Media
has been completed, press OK to continue.
case of BT05 SW, Removable Media) and selecting VERIFY as shown in Figure 4-31.
Figure 4-32 Format Successful Pop-up Menu