4 Using the Measurement Functions

Scope Marker

Scope Horizontal

Scope Preset

118
Select the Markers button to configure the measurements that
appear on the bottom of the display in Marker view. Measurements
apply to the portion of the waveform between the two markers. You
can only select a maximum of five measurements to be displayed.
Select the Horizontal button to configure the horizontal properties.
The Sample Points field lets you specify the number of points in a
scope trace. The maximum number of points that can be specified
depends on the number of scope traces that have been enabled. The
minimum number of points that can be specified is 1024.
1 trace enabled: 256 K points
2 traces enabled: 128 K points
The Power trace counts as 2 traces, since voltage and current must be
measured to calculate power. If the voltage and current traces have
already been selected, the Power trace is not counted.
The Horizontal Offset Reference places the reference point on the
left, right, or center of the display. This is the position of the trigger if
no offset has been set. Left lets you see the waveform after the trigger
event. Center lets you see the waveform before and after the trigger
event. Right lets you see the waveform up to the trigger event.
Select the Preset button to return the Scope View to the power-on
display settings. The vertical offset of each trace will be set to a
different value. This is to prevent the traces from overlapping. The
offset is referenced to the horizontal center line of the grid.
4 traces enabled: 64 K points
8 traces enabled: 32 K points
(maximum points = 256 K/# of traces)
Model N6705 User's Guide