SCPI Command Summary

NOTE
SCPI Command
ABORt
:ACQuire (@chanlist)
:DLOG
:ELOG (@chanlist)
:HISTogram (@chanlist)
:TRANsient (@chanlist)
CALibrate
:CURRent
[:LEVel] <NRf>, (@channel)
:LIMit
:NEGative <NRf>, (@channel)
:POSitive <NRf>, (@channel)
:MEASure <NRf>, (@channel)
:PEAK (@channel)
:DATA <NRf>
:DATE <"date">, (@channel)
:DPRog (@channel)
:LEVel P1 | P2 | P3
:PASSword <NRf>
:RESistance 20| 6, (@channel)
:SAVE
:STATE <Bool> [,<NRf>]
:VOLTage
[:LEVel] <NRf>, (@channel)
:CMRR (@channel)
:LIMit
:POSitive <NRf>, (@channel)
:MEASure <NRf>, (@channel)
:AUXiliary (@channel)
DISPlay
[:WINDow]
:VIEW METER1 | METER4
FETCh
[:SCALar]
:CURRent
[:DC]? (@chanlist)
:ACDC? (@chanlist)
:HIGH? (@chanlist)
:LOW? (@chanlist)
:MAXimum? (@chanlist)
:MINimum? (@chanlist)
:POWer
[:DC]? (@chanlist)
184
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. Not all commands apply to all models.
Description
Resets the measurement trigger system to the Idle state
Stops the internal data logger
Stops the external data logger
Stops the histogram current measurement
Resets the transient trigger system to the Idle state
Calibrates the output current programming
Calibrates the negative current limit (only on N6783A)
Calibrates the positive current limit (only on N678xA SMU)
Calibrates the current measurement
Calibrates the peak current limit (only on N675xA, N676xA)
Enters the calibration value
Sets the calibration date
Calibrates the current downprogrammer
Advances to the next calibration step
Sets the numeric calibration password
Calibrates the output resistance (only on N6781A)
Saves the new cal constants in non-volatile memory
Enables/disables calibration mode
Calibrates the output voltage programming
Calibrates common mode rejection ratio (only N675xA, N676xA)
Calibrates the positive voltage limit (only on N678xA SMU)
Calibrates the voltage measurement
Calibrates the auxiliary voltage measurement (only on N6781A)
Selects 1-channel or 4-channel meter view
Returns the average current
Returns the total rms current (AC + DC)
Returns the high level of a current pulse
Returns the low level of a current pulse
Returns the maximum current
Returns the minimum current
Returns the average output power