Agilent Technologies N6705 User Manual page 96

Dc power analyzer
Table of Contents

Advertisement

3 Using the Source Functions
96
User-Defined Parameters
Parameter:
Step <n>
Current
Voltage
or
Time
Trigger
Add
Delete
Clear
Import (.csv format)
Export (.csv format)
From the remote interface:
The current or voltage value of five steps:
ARB:CURR:UDEF:LEV 1,2,3,4,5,(@1)
ARB:VOLT:UDEF:LEV 1,2,3,4,5,(@1)
The time that the output stays at the steps:
ARB:CURR:UDEF:DWEL 1,2,3,2,1,(@1)
ARB:VOLT:UDEF:DWEL 1,2,3,2,1,(@1)
Generate an external trigger signal at the start of the step
(the trigger is generated at the start of step #3):
ARB:CURR:UDEF:BOST 0,0,1,0,0,(@1)
ARB:VOLT:UDEF:BOST 0,0,1,0,0,(@1)
Description:
Each step of the waveform includes a voltage or
current parameter, dwell time, and trigger option.
The total number of steps determines the length.
Use the ▲ ▼ navigation keys to scroll through the steps.
The current or voltage value of the step.
The time that the output stays at the step.
Generate an external trigger signal at the start of
the step when checked.
Inserts a step below the selected step; values are
copied from the previous step.
Deletes the presently selected step.
Clears all steps.
Imports a current or voltage Arb list.
Exports a voltage or current Arb list.
Model N6705 User's Guide