2 Installation
1. To front
panel
binding post.
2. 20 A
connector
1. To front
panel
binding post.
2. Aux Meas
connector
3. To rear
panel
connector
A. Adapter A
required for
N6705A
mainframes
1
only
30
Step 3. Connect the front panel wire harness.
For power modules that use 12 A output connectors –
Simply push the 12 A connector plug into the power module. Tighten
the locking screws on the connector.
For power modules with 20 A output connectors –
On N6705B mainframes, remove the 12 A connector plug from the
wire harness and install the 20 A connector plug provided with the
power module. Observe the output color code. Tighten all connector
screws. Install the connector in the module.
On N6705A mainframes, install the wire harness with the 20 A
connector plug directly into the power module.
For power modules with 50 A output connectors –
see "High Current Output Connections".
1
For Agilent N678xA SMU power modules –
Remove the 12A connector plug from the wire harness and install the
8-pin connector plug provided with the power module. Install the
front panel cable wires in the output connector as shown. Observe
the output color code. Tighten all connector screws. An adapter "A"
is required to install modules in Agilent N6705A mainframes.
For model N6781A, install the auxiliary measurement (Aux Meas)
cable. Remove the Aux Meas cable from its storage location and
insert the connector into the power module. The color code
corresponds to the rear panel Aux Voltage Measurement label.
Aux Meas +
Aux Meas –
– –
+ +
G
+S G –S
1
3
2
+
+
S
S
+
+s +
-s
2
+
S
S
Model N6705 User's Guide