Agilent Technologies N6705 User Manual page 188

Dc power analyzer
Table of Contents

Advertisement

Appendix B SCPI Commands
SCPI Command
SENSe:HISTogram (continued)
:FUNCtion
:CURRent <Bool>, (@chanlist)
:SWEep
:OFFSet
:POINts <NRf+>, (@chanlist)
:POINts <NRf+>, (@chanlist)
:TINTerval <NRf+>, (@chanlist)
:RESolution RES20 | RES40, (@chanlist)
:VOLTage
[:DC]:RANGe
[:UPPer] <NRf+>, (@chanlist)
:AUTO <Bool>, (@chanlist)
:WINDow
[:TYPE] HANNing | RECTangular, (@chanlist)
[SOURce:]ARB
:COUNt <NRf+> | INFinity, (@chanlist)
:CURRent | :VOLTage
:CDWell
[:LEVel] < NRf+> {,<NRf>}, (@chanlist)
:DWELl < NRf+>, (@chanlist)
:POINts? (@<chanlist>)
:CONVert (@chanlist)
:EXPonential
:END
[:LEVel] < NRf+>, (@chanlist)
:STARt
[:LEVel] < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:TCONstant < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:PULSe
:END
:TIMe < NRf+>, (@chanlist)
:STARt
[:LEVel] < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:TOP
[:LEVel] < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:RAMP
:END
[:LEVel] < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:RTIMe < NRf+>, (@chanlist)
:STARt
[:LEVel] < NRf+>, (@chanlist)
:TIMe < NRf+>, (@chanlist)
:SINusoid
:AMPLitude < NRf+>, (@chanlist)
:FREQuency < NRf+>, (@chanlist)
:OFFSet < NRf+>, (@chanlist)
188
Description
Enables/disables current histogram (only on N6781A, N6782A)
Defines the trigger offset in the measurement sweep
Defines the number of data points in the measurement
Sets the measurement sample interval
Sets the measurement resolution
Selects the voltage measurement range
Enables/disables measurement autoranging (N6781A, N6782A)
Selects the window type
Sets the Arb repeat count
Specifies current or voltage Arbs
Sets the list of the constant dwell Arb
Sets the dwell time for the constant dwell Arb
Returns the number of constant-dwell Arb points
Converts the selected Arb to a user-defined list
Sets the end level of the exponential Arb
Sets the initial level of the exponential Arb
Sets the length of the start time or delay
Sets the time constant of the exponential Arb
Sets the time of the exponential Arb
Sets the length of the end time
Sets the initial level of the pulse
Sets the length of the start time or delay
Sets the top level of the pulse
Sets the length of the pulse
Sets the end level of the ramp
Sets the length of the end time
Sets the rise time of the ramp
Sets the initial level of the ramp
Sets the length of the start time or delay
Sets the amplitude of the sine wave
Sets the frequency of the sine wave
Sets the DC offset of the sine wave
Model N6705 User's Guide

Advertisement

Table of Contents
loading

Table of Contents