Model N6705 User's Guide
Exponential Parameters
Parameter:
Start Setting (I
or V
)
0
0
End Setting (I
or V
)
1
1
Delay (T
)
0
Time (T
)
1
Time Constant (T
)
C
From the remote interface:
The setting before the waveform:
ARB:CURR:EXP:STAR 0,(@1)
ARB:VOLT:EXP:STAR 0,(@1)
The end setting of the waveform:
ARB:CURR:EXP:END 1,(@1)
ARB:VOLT:EXP:END 5,(@1)
The delay after the trigger is received but before the waveform starts:
ARB:CURR:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
Time for the amplitude to go from the start setting to the end setting:
ARB:CURR:EXP:TIM 0.75,(@1)
ARB:VOLT:EXP:TIM 0.75,(@1)
The time constant of the curve:
ARB:CURR:EXP:TCON 0.1,(@1)
ARB:VOLT:EXP:TCON 0.1,(@1)
Using the Source Functions 3
Description:
The setting before the waveform.
The end setting of the waveform.
The delay after the trigger is received
.
before the waveform starts
Time for the amplitude to go from the start
setting to the end setting.
The time constant of the curve.
but
95