Agilent Technologies N6705 User Manual page 125

Dc power analyzer
Table of Contents

Advertisement

Symbol/Field:
1
m1/m2 points
2
Delta
3
m2
4
m1
5
Min
6
Avg
7
Max
8
V p-p
Ah (if selected)
Wh (if selected)
Model N6705 User's Guide
Marker View
Description:
Shows where the measurement markers intersect the selected waveform.
Data values at the bottom of the display are referenced to the intersect
locations of the markers. Calculations are based on the data points in
between the intersect locations.
Indicates the delta or absolute difference between the markers in units
(volts, amps, or watts) and in time (seconds).
Indicates the m2 marker value in volts, amps, or watts at the intersection
point. Also indicates the distance in time that the m2 marker is in relation to
the present trigger position.
Indicates the m1 marker value in volts, amps, or watts at the intersection
point. Also indicates the distance in time of the m1 marker in relation to the
present trigger position.
Indicates the minimum data value (in volts, amps, or watts) between the
marker locations of the selected waveform. Also indicates the distance in
time of the minimum value in relation to the present trigger position.
Calculates the average data value (in volts, amps, or watts) between the
marker locations of the selected waveform. Time indicates the time between
markers over which the average value is calculated.
Indicates the maximum data value (in volts, amps, or watts) between the
marker locations of the selected waveform. Also indicates the distance in
time of the maximum value in relation to the present trigger position.
Calculates the difference between the maximum and minimum values. Time
information is not valid for calculated p-p values.
Calculates the Amp-hours between the marker locations. To view the Amp
hours, you must unselect one of the other measurements in the Datalogger
Marker Properties window. Only 5 measurements may be displayed at a time.
Calculates the Watt-hours between the marker locations. To view the Watt
hours, you must unselect one of the other measurements in the Datalogger
Marker Properties window. Only 5 measurements may be displayed at a time.
Using the Measurement Functions 4
125