Data Logger Preset - Agilent Technologies N6705 User Manual

Dc power analyzer
Table of Contents

Advertisement

4 Using the Measurement Functions

Data Logger Preset

132
From the remote interface:
The following commands position the markers. To position the two
data log markers at 100 seconds and at 200 seconds from the start
trigger on the data log:
SENS:DLOG:MARK1 100
SENS:DLOG:MARK2 200
The following commands return data between the two markers. To
return the average current or voltage between the markers:
FETC:DLOG:CURR? (@1)
FETC:DLOG:VOLT? (@1)
To return the minimum current or voltage between the markers:
FETC:DLOG:CURR:MIN? (@1)
FETC:DLOG:VOLT:MIN? (@1)
To return the maximum current or voltage between the markers:
FETC:DLOG:CURR:MAX? (@1)
FETC:DLOG:VOLT:MAX? (@1)
To return the peak-to-peak current or voltage between the markers:
FETC:DLOG:CURR:PTP? (@1)
FETC:DLOG:VOLT:PTP? (@1)
Select the Preset button to return the Data Logger View to the power-
on display settings. In the power-on settings, the vertical offset of
each trace is set to a different value. This is to prevent the traces
from overlapping. The offset is referenced to the horizontal center
line of the grid.
Model N6705 User's Guide