Agilent Technologies N6705 User Manual page 93

Dc power analyzer
Table of Contents

Advertisement

Model N6705 User's Guide
Pulse Parameters
Parameter:
Start Setting (I
or V
)
0
0
Pulse Setting (I
or V
)
1
1
Delay (T
)
0
Pulse Width (T
)
1
End Time (T
)
2
From the remote interface:
The setting before and after the pulse:
ARB:CURR:PULS:STAR 0,(@1)
ARB:VOLT:PULS:STAR 0,(@1)
The amplitude of the pulse:
ARB:CURR:PULS:TOP 1,(@1)
ARB:VOLT:PULS:TOP 10,(@1)
The delay after the trigger is received but before the pulse starts:
ARB:CURR:PULS:STAR:TIM 0.25,(@1)
ARB:VOLT:PULS:STAR:TIM 0.25,(@1)
The width of the pulse:
ARB:CURR:PULS:TOP:TIM 0.5,(@1)
ARB:VOLT:PULS:TOP:TIM 0.5,(@1)
The time the output remains at the end setting after the pulse:
ARB:CURR:PULS:END:TIM 0.25,(@1)
ARB:VOLT:PULS:END:TIM 0.25,(@1)
Using the Source Functions 3
Description:
The setting before and after the pulse.
The amplitude of the pulse.
The delay after the trigger is received but
before the pulse starts.
The width of the pulse.
The time the output remains at the end
setting after the pulse completes.
93