3 Using the Source Functions
80
Step 4 – Select the trigger source:
To specify a Trigger source for the arbitrary waveforms, press the
Arb key, then select the Trigger Source field. The same trigger
source will be used to trigger all of the arbitrary waveforms.
Arb Run/Stop key selects the front panel Arb Run/Stop key as the
trigger source. BNC Trigger In selects the rear trigger input BNC
connector as the trigger source. Remote command selects a remote
interface command as the trigger source.
Step 4 – Preview and run the Arb:
The Arb Preview dialog shown above provides a preview of the User-
defined waveform that will be run on output 1.
Select Meter View or Scope View to display the Arb.
Press the Output 1 On key to enable the output.
Press the Arb Run/Stop key to run the Arb.
From the remote interface:
The following commands select, program, and trigger a User-defined
voltage waveform of five steps on output 1:
ARB:FUNC:TYPE VOLT,(@1)
ARB:FUNC:SHAP UDEF,(@1)
ARB:VOLT:UDEF:LEV 10,20,30,40,50,(@1)
ARB:VOLT:UDEF:DWEL 1,2,3,2,1,(@1)
ARB:VOLT:UDEF:BOST 0,0,1,0,0,(@1)
ARB:TERM:LAST OFF,(@1)
To set up the transient trigger system and trigger the Arb:
VOLT:MODE ARB,(@1)
TRIG:ARB:SOUR BUS
OUTP ON,(@1)
INIT:TRAN(@1)
*TRG
Model N6705 User's Guide