Agilent Technologies N6705 User Manual page 111

Dc power analyzer
Table of Contents

Advertisement

Model N6705 User's Guide
Step 5 – Turn on the outputs and measure the voltage:
Press Scope View to return the display the scope view.
• Press the Run/Stop key to run the scope. When this key is lit, it
indicates that the scope is running.
• Press the All Outputs On key to start the output sequence and
trigger the scope.
You should see the output waveforms displayed as follows:
Note that the delay for output 1 does not start until 57 milliseconds
have elapsed. This is because the internal delay for the mainframe
used in this example is 57 milliseconds. This built in delay precedes
any user-defined delays that have been programmed.
Also note that outputs 3 and 4 start at the specified delay time but do
not ramp up as fast as outputs 1 and 2. This is because Outputs 1
and 2 are "Precision" and "High Performance" modules, while
outputs 3 and 4 are "DC Power" modules, which have a slower turn-
on ramp. Refer to chapter 1 under "Power Module Features".
From the remote interface:
You cannot program the scope from the remote interface.
Using the Measurement Functions 4
111