Agilent Technologies N6705 User Manual page 115

Dc power analyzer
Table of Contents

Advertisement

Knob:
1
Vertical Volts/Div
2
Vertical Offset
3
Trigger Level
4
Horizontal Time/Div
5
Horizontal Offset
6
Marker 1/Marker 2
Model N6705 User's Guide
Using the Waveform Display Knobs
1
2
Description:
Makes the waveform bigger or smaller vertically in relation to its ground
reference. Specified in volts/division or amps/division on the y axis. On
outputs with multiple ranges, when Knob Control is selected in the Scope
Range Property window, adjusting the vertical gain will automatically select
lower measurement ranges for better resolution. If the vertical gain causes
the trace to be out of view, arrow symbols
will indicate the direction of the trace.
Moves the ground reference of the trace up or down in relation to the
horizontal center line of the grid. The offset popup that
appears in the upper right corner of the display shows
how far the ground reference of the selected trace is above or below the
horizontal center line of the grid.
Positive values indicate that the center line is above the ground reference.
Negative values indicate that the center line is below the ground reference.
Moves the trigger level up and down when a voltage or current level is the
trigger source. The trigger level is identified by the
level is out of view, an arrow
Push the Trigger Level knob to auto-scale the traces on the display. This will
trigger a new measurement.
Stretches or shrinks the waveform horizontally around the horizontal offset
reference. Specified in time/division on the x axis. Applies to ALL traces.
Moves the waveform to the right or left of the horizontal offset reference.
The trigger point is indicated by the solid arrow.
Moves the measurement markers right or left. Press Scope View to display
the markers. Markers are displayed on the selected trace. Values at the
bottom of the display are referenced to the intersection of the markers. If a
marker is out of view, an arrow indicates the direction of the marker.
Push the Marker 1 or Marker 2 knobs to reset the
markers. When pushed, the following menu appears.
Push the Enter key to Reset the marker. Push Enter
again to undo the reset action. Scroll down and
select Marker options to access the Scope Marker Properties window. Scroll
down and select Jump to peak to move the marker to the peak measurement
point of the trace.
Using the Measurement Functions 4
symbol. If the trigger
indicates the direction of the trigger level.
3
4
5
6
115