From the remote interface:
To program turn-on and turn-off delays for channels 1 through 4:
OUTP:DEL:RISE .01,(@1)
OUTP:DEL:RISE .02,(@2)
OUTP:DEL:RISE .03,(@3)
OUTP:DEL:RISE .04,(@4)
OUTP:DEL:FALL .04,(@1)
OUTP:DEL:FALL .03,(@2)
OUTP:DEL:FALL .02,(@3)
OUTP:DEL:FALL .01,(@4)
To only include outputs 1 and 2 in a sequence and specify a different
delay offset:
OUTP:COUP:CHAN 1,2
OUTP:COUP:DOFF:MODE MAN
OUTP:COUP:DOFF .050
To query the delay offset of the slowest power module in the
mainframe (the maximum delay offset) in seconds:
OUTP:COUP:MAX:DOFF?
To turn on two coupled outputs in a sequence:
OUTP ON,(@1:2)
Using the Source Functions 3