Using the Data Logger Functions

NOTE

Logging Data

Model N6705 User's Guide
The Data Logger function is not available if Option 055 has been ordered.
The Data Logger is similar to the Scope View function except that it
lets you view and log output voltage and current data for up to
99,999 hours.
As in Scope View, you can configure the Data Logger View to display
voltage or current waveforms for all outputs. Power waveforms can
be displayed for all outputs because of the interleaved data logging
capability. See "Data Logger Sampling Modes" later in this chapter.
The display functions like a strip chart recorder. Use the Waveform
Display knobs to scroll through the data. Unless specified otherwise,
data is automatically stored to a file named default.dlog.
In the following data log example, a user-defined arbitrary waveform
is captured on the data logger. The data logger records the actual
output voltage of the arbitrary waveform.
Step 1 – Program the arbitrary waveform:
Configure the user-defined arbitrary waveform as described in
chapter 3 under "Configuring User-Defined Arbs". Program the
voltage and time values as follows:
• Step 0: 10 V; 1 s
• Step 1: 20 V; 1 s
• Step 2: 30 V; 1 s
• Step 3: 40 V; 1 s
• Step 4: 50 V; 1 s
• Repeat Count: 5
Using the Measurement Functions 4
119