Model N6705 User's Guide
Step 3 – Enable the output:
Press the color-coded On key to enable an individual output. When
an output is on, the On key for that output is lit. When an output is
Off, the On key is not lit. The All Outputs On and Off keys turn all
outputs on or off simultaneously.
The state of a disabled output (output off) is a condition of zero
output voltage and zero source current.
The red Emergency Stop key turns all outputs off immediately without any
NOTE
output off delays.
Step 4 – View the output voltage and current:
Select Meter View to view the output voltage and current. When an
output is enabled, the front panel meters continuously measure and
display the output voltage and current.
From the remote interface:
A channel parameter is required with each SCPI command to select
an output. For example, (@1) selects output 1, (@2,4) selects output 2
and 4, and (@1:4) selects outputs 1 through 4. The output list must be
preceded with an @ symbol and be enclosed in parentheses ().
To set only output 1 to 10.02 V and 1 A:
VOLT 10.02,(@1)
CURR 1,(@1)
To set the output current limit to 1 A for Agilent Models N678xA SMU
and N6783A:
CURR:LIM 1,(@1)
To set the output voltage of all outputs to 10 V:
VOLT 10.02,(@1:4)
To enable only output 1:
OUTP ON,(@1)
To enable output 1 and output 3:
OUTP ON,(@1,3)
To measure the output voltage and current of output 1:
MEAS:VOLT? (@1)
MEAS:CURR? (@1)
Using the Source Functions 3
61