86
Under Path/Filename, click Browse and specify the location where
the import file is located or where the export file will be placed.
Internal:\ specifies the instrument's internal memory. External:\
specifies the Memory port on the front panel.
Select Load to load the file. Select Save to save the file.
Step 4 – Select the trigger source:
To specify a Trigger source for the arbitrary waveforms, press the
Arb key, then select the Trigger Source field. The same trigger
source will be used to trigger all of the arbitrary waveforms.
Arb Run/Stop key selects the front panel Arb Run/Stop key as the
trigger source. This means that the Arb will be triggered as soon as
you run it. BNC Trigger In selects the rear trigger input BNC
connector as the trigger source. Remote command selects a remote
interface command as the trigger source.
Step 4 – Preview and run the Arb:
The Arb Preview dialog shown above provides a preview of the Arb
Sequence that will be run on output 1.
Select Meter View or Scope View to display the Arb.
Press the Output 1 On key to enable the output.
Press the Arb Run/Stop key to run the Arb.
From the remote interface:
Observe the following when creating or editing an Arb sequence:
The Arb function type (voltage or current) must match the
Arb type specified in each sequence step.
Sequence steps must be specified sequentially. The last value
in the parameter list is the sequence step number.
Model N6705 User's Guide