Agilent Technologies N6705 User Manual page 97

Dc power analyzer
Table of Contents

Advertisement

Model N6705 User's Guide
Constant-Dwell Parameters
Parameter:
Dwell per Step
Import (.csv format)
Export (.csv format)
Min
Max
Avg
Points
Time
Repeat Count
From the remote interface:
The dwell time for each step in seconds.:
ARB:CURR:CDW:DWEL 0.01,(@1)
ARB:VOLT:CDW:DWEL 0.01,(@1)
The current or voltage value of ten steps:
ARB:CURR:CDW 0.5,1,1.5,2,2.5,3,3.5,4,4.5,5,(@1)
ARB:VOLT:CDW 1,2,3,4,5,6,7,8,9,10,(@1)
Using the Source Functions 3
Description:
The dwell time for each step in seconds.
Values range from10.24 μs to 0.30 s.
Imports a current or voltage CD Arb list.
Exports a current or voltage CD Arb list.
The minimum value of the imported Arb
The maximum value of the imported Arb
The average value of the imported Arb
The number of points in the imported Arb
The total time of the imported Arb
The maximum repeat count for CD Arbs is 256.
97

Advertisement

Table of Contents
loading