3 Using the Source Functions
90
Ramp Parameters
Parameter:
Start Setting (I
or V
)
0
0
End Setting (I
or V
)
1
1
Delay (T
)
0
Ramp Time (T
)
1
End Time (T
)
2
From the remote interface:
The setting before the ramp:
ARB:CURR:RAMP:STAR 0,(@1)
ARB:VOLT:RAMP:STAR 0,(@1)
The setting after the ramp:
ARB:CURR:RAMP:END 1,(@1)
ARB:VOLT:RAMP:END 5,(@1)
The delay after the trigger is received but before the ramp starts:
ARB:CURR:RAMP:STAR:TIM 0.25,(@1)
ARB:VOLT:RAMP:STAR:TIM 0.25,(@1)
The time that the output ramps up:
ARB:CURR:RAMP:RTIM 0.5,(@1)
ARB:VOLT:RAMP:RTIM 0.5,(@1)
The time the output remains at the end setting after the ramp:
ARB:CURR:RAMP:END:TIM 0.01,(@1)
ARB:VOLT:RAMP:END:TIM 0.01,(@1)
Description:
The setting before the ramp.
The setting after the ramp.
The delay after the trigger is received but
before the ramp starts.
The time that the output ramps up.
The time the output remains at the end
setting after the ramp completes.