Table of Contents

Advertisement

3 Using the Source Functions
82
Specify what happens when the waveform completes – select
whether the output returns to the DC value that was in effect before
the waveform started, or whether the output should remain at the
last Arb value.
Specify if the Arb should repeat, either continuously or only for a
specified repeat count. A repeat count of 1 runs the Arb once.
Step 3 – Importing or exporting Constant-dwell Arb data
Refer to "Importing and Exporting User-Defined and Constant-Dwell
Arb Data" later in this chapter for information on how to create .csv
current or voltage data files.
Select the Arb.csv file type format. When importing, specify the
output on which the Constant-dwell Arb will run. When exporting,
specify the output from which you want to export the Arb.
Under Path/Filename, click Browse and specify the location where
the import file is located or where the export file will be placed.
Internal:\ specifies the instrument's internal memory. External:\
specifies the Memory port on the front panel.
Select Import to import the file. Select Export to export the file.
Step 4 – Select the trigger source:
To specify a Trigger source for the arbitrary waveforms, press the
Arb key, then select the Trigger Source field. The same trigger
source will be used to trigger all of the arbitrary waveforms.
Arb Run/Stop key selects the front panel Arb Run/Stop key as the
trigger source. BNC Trigger In selects the rear trigger input BNC
connector as the trigger source. Remote command selects a remote
interface command as the trigger source.
Model N6705 User's Guide