Single; Auto - Rigol DS6000 Series User Manual

Ds6000 series digital oscilloscope
Hide thumbs Also See for DS6000 Series:
Table of Contents

Advertisement

RIGOL

Single

Auto

1-20
Press this key to set the trigger mode to "Single". In
single trigger mode, the oscilloscope generates a trigger
when the trigger conditions are met and then stops.
Press this key to turn on the waveform auto setting
function. The oscilloscope will automatically adjust the
vertical scale, horizontal time base and trigger mode
according to the input signal to realize optimum
waveform display and provide quick parameter
measurement (please refer to Quick Measurement
after Auto).
Note: auto setting requires that the frequency of the
signal under test should be no lower than 50 Hz, the duty
cycle be greater than 1% and the amplitude be at least
20 mVpp. If the parameters exceed these limits, "Auto
failed!" would be displayed after pressing this key and the
quick parameter measurement menu might not be
displayed.
1 Quick Start
DS6000 User's Guide