Manual Mode - Rigol DS6000 Series User Manual

Ds6000 series digital oscilloscope
Hide thumbs Also See for DS6000 Series:
Table of Contents

Advertisement

RIGOL

Manual Mode

In this mode, a pair of cursors will appear. You can adjust the cursors to measure the
X (or Y), X increment (or Y increment), and reciprocal of the X increment on the
selected source (CH1-CH4 or MATH).
Press CURS  Mode  "Manual" to turn the manual cursor function on and the
measurement results will be displayed at the upper left corner of the screen in the
following mode.
The X or Y value at cursor A (CurA): X value takes the trigger position as
reference and Y value takes the channel GND as reference.
The X or Y value at cursor B (CurB): X value takes the trigger position as
The horizontal or vertical difference between cursor A and B (△X or △Y).
reference and Y value takes the channel GND as reference.
The reciprocal of the horizontal difference between cursor A and B (1/△X).
If needed, please refer to the following steps to modify the parameters of manual
cursor measurement.
1.
Select Cursor Type
Press Type to select "Time" or "Voltage". The time cursor is a pair of vertical
dotted lines and is usually used to measure time parameters. The voltage cursor
is a pair of horizontal dotted lines and is usually used to measure voltage
parameters.
2.
Select Measurement Source
Press Source to select the waveform in the analog channels (CH1-CH4) or math
operation results (MATH) for measurement (only channels enabled can be
selected). If "None" is selected, no cursor will be displayed.
6-28
6 To Make Measurements
DS6000 User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds6104Ds6102Ds6064Ds6062

Table of Contents