1. Press SAVE  Factory to restore the oscilloscope to its default configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" under the
probe compensation signal output terminal
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
4. Press Auto.
5. Observe the waveform on the display. In normal condition, the display should be
a square waveform as shown in the figure below:
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation" in the next section.
1-10
Figure 1-8 To Use the Compensation Signal
Figure 1-9 Square Waveform
1 Quick Start
Compensation Signal
Output Terminal
Ground Terminal
DS6000 User's Guide