Rigol DS6000 Series User Manual page 131

Ds6000 series digital oscilloscope
Hide thumbs Also See for DS6000 Series:
Table of Contents

Advertisement

6 To Make Measurements
4.
Set the Vertical Scale
In FFT measurement, the unit of the horizontal axis changes from time to
frequency. Use HORIZONTAL
horizontal axis respectively.
The unit of the vertical axis can be dB or Vrms which display the vertical
amplitude in logarithmic mode and linear mode respectively. If you need to
display the FFT frequency spectrum in a relatively larger dynamic range, dB is
recommended. Press Scale to select the desired unit and the default is Vrms.
Use VERTICAL
amplitude and position of the FFT frequency spectrum respectively.
Tips
Signals with DC components or deviation would cause error or deviation of
the FFT waveform components. To reduce the DC components, set the
"Channel Coupling" to "AC".
To reduce the random noise and aliasing frequency components of
repetitive or single pulse, set the "Acquisition Mode" of the oscilloscope
to "Average".
DS6000 User's Guide
SCALE to set the scale and position of the
SCALE and HORIZONTAL
SCALE to set the vertical
RIGOL
6-7