Rear Panel Overview

1. Trig Out/Calibration
This connector can output various signals (press UTIL  Aux Output to select
the desired output type.):
1) TrigOut: after this type is selected, the oscilloscope output a signal that can
reflect the current capture rate of the oscilloscope at each trigger.
2) Fast: after this type is selected, the oscilloscope can output a fast edge
signal which can be used in the self-calibration of the oscilloscope.
3) GND: after this type is selected, the oscilloscope can output a ground level.
4) PassFail: after this type is selected, the oscilloscope will output a pulse
signal when failed waveforms are detected. Connect this signal to other
control systems to conveniently view the test results.
2. Reference Clock
Provide more precise sample clock signal for the oscilloscope and it can also
1-14