Select Measurement Source
Press CursorA to select the waveform in analog channels (CH1-CH4) or math
operation results (MATH) as the measurement source of cursor A (only channels
enabled are available). You can also select "None", namely do not use cursor A.
Press CursorB to select the waveform in analog channels (CH1-CH4) or math
operation results (MATH) as the measurement source of cursor B (only channels
enabled are available). You can also select "None", namely do not use cursor B.
2.
Adjust Cursor Position (note that under the same menu page, you can
also press
Adjust cursor A: press CursorA and use
A. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor B: press CursorB and use
B. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor A and B at the same time: press CursorAB and use
adjust the positions of cursor A and B at the same time. During the
adjustment, the measurement results will change accordingly. The
adjustable range is limited within the screen.
Note: the vertical cursor will track the marked point (namely jumps up and
down with the transient moving of the waveform). Thus, the Y value might
change even though you do not adjust the cursor.
3.
Measurement Example
Use cursor A and B to measure the waveforms in CH1 and CH2 respectively. You
would discover that the cursors would track the marked points when the
waveforms are expanded or compressed horizontally.
6-32
continuously to switch the current cursor)
6 To Make Measurements
to adjust the position of cursor
to adjust the position of cursor
DS6000 User's Guide