RIGOL
Voltage Parameters
Vmax
Vtop
Vpp
Vbase
Vmin
1.
Vmax: the voltage value from the highest point of the waveform to the GND.
2.
Vmin: the voltage value from the lowest point of the waveform to the GND.
3.
Vpp: the voltage value from the highest point to the lowest point of the
waveform.
4.
Vtop: the voltage value from the flat top of the waveform to the GND.
5.
Vbase: the voltage value from the flat base of the waveform to the GND.
6.
Vamp: the voltage value from the top of the waveform to the base of the
waveform.
7.
Vavg: the arithmetic average value on the whole waveform or on the gating
area.
Average
=
n
points being measured.
8.
Vrms: the root mean square value on the whole waveform or the gating area.
n
x
RMS
=
i
=
1
n
points being measured.
6-18
Overshoot
Vamp
Preshoot
x
i
, where,
x is the i
i
2
i
, where,
x is the i
i
th
point being measured, n is the number of
th
point being measured, n is the number of