2 To Set the Vertical System
Table 2-1 Probe Attenuation Coefficient
Menu
0.01X
0.02X
0.05X
0.1X
0.2X
0.5X
1X
2X
5X
10X
20 X
50 X
100 X
200 X
500 X
1000 X
When "DiffProbe" is inserted, the oscilloscope recognizes it automatically and the
following menus are added.
3. Front-End
RP7150 active probe provides "Single-end" and "Difference" probe heads. Press
this softkey to select the desired probe head.
4. Probe-Cal
Connect the differential probe to the channel input terminal (such as CH1) of the
oscilloscope correctly, then connect the fast edge signal output from the [Trig
Out/Calibration] connector at the rear panel of the instrument to the probe
and CH2 respectively. At this point, press this softkey and the oscilloscope starts
to calibrate the probe.
Note: if the probe is connected to CH1, CH3 or CH4, the fast edge signal should
be connected to CH2; if the probe is connected to CH2, the fast edge signal
should be connected to CH1.
5. Probe Info
Press this softkey to view various information about the probe, such as the
manufacturer, model, serial number and the date of last calibration.
DS6000 User's Guide
Attenuation coefficient
1:100
1:50
1:20
1:10
1:5
1:2
1:1
2:1
5:1
10:1
20:1
50:1
100:1
200:1
500:1
1000:1
RIGOL
2-5