3 To Set the Horizontal System
Note:
In Y-T mode, the oscilloscope could use arbitrary sample rate (within the
guaranteed range) to capture waveform. The maximum sample rate of Y-T
mode is 2.5 GSa/s. Generally, reducing the sample rate properly could get a
Lissajous figure with relatively better display effect.
When X-Y mode is turned on, "Delayed" will be turned off automatically.
The following functions are not available in X-Y mode:
Auto measure, cursor measure, math operation, delayed sweep, vector display
type, HORIZONTAL
acquisition mode, Pass/Fail test and waveform recording.
Method 2: Use the shortcut measurement function
Please refer to "Phase A→B " and "Phase A→B " measurement functions of "Delay
and Phase" on page 6-17.
DS6000 User's Guide
POSITION, trigger control, memory depth,
RIGOL
3-7